Animal ventilator is a commonly used experimental equipment, widely used in scientific research experiments such as basic medicine, clinical medicine, and animal medicine for artificial respiration, respiratory management, animal first aid, respiratory therapy, etc. Its control is accurate, convenient and practical, does not require high-pressure air source, accurate tidal volume output, and stable performance.
Technical Parameter
1. Instrument size: 300x240x100mm
2. Respiratory rate regulation range: 20-60 beats per minute
3. Tidal volume adjustment range: 0.2~90ml
4. Inhalation to respiration ratio adjustment: 1:9~9:1 adjustable
5. Airway pressure waveform display range: 7Kpa
6. Automatic pressure relief protection when airway pressure exceeds 6Kpa
7. Ventilator interface specification: 4mm diameter 304 stainless steel connector
8. Parameter display and operation: 5-inch touch LCD screen display
9. Screen size: length 290mm, width 210mm, height 100mm
10. Screen resolution: 800x480
11. Input voltage: AC220V
12. Noise: ≤ 67dB
13. Air or non flammable mixed gas
14. Scope of application: Suitable for rabbits and cats
Reference Parameter Settings
Animal |
Weight(kg) |
Respiratory rate (times/min) |
Tidal volume (ml) |
Macaque |
2-4 |
30-50 |
20-100 |
Dog |
10 |
16-20 |
150-200 |
Cat |
2-3 |
20-30 |
20-70 |
Rabbit |
2-3 |
35-50 |
20-70 |
Guinea pig |
0.27-0.94 |
70-110 |
4-25 |
Rat |
0.15-0.40 |
70-110 |
4-25 |
Mice |
0.02-0.04 |
100-130 |
1-3 |
Selection Guide
M5209S Small animal ventilator |
M5209-1 Small animal ventilator - compatible with cats, dogs, and rabbits |
M5209-2 Rabbits and cats using ventilators |
M5209 Large animal ventilator |