In
animal experiments, it is often necessary to perform microinjection on
animals. An automatic microinjection pump is used, which is convenient
to use. The syringe of the injector is installed on the injection pump
and connected to the gripper through a conduit. The precise movement of
the injector piston is controlled by a microcomputer to drive the unit,
achieving injection or recovery of liquids in experimental animals and
leaving more operating space for the experiment.
M5262 four channel
independent control micro injection pump, high-definition touch screen
operation, Ditronstep micro step control system, used for precise
control and delivery of micro liquid drugs or solutions. One to four
structure, independent control of four channels; high-precision; Low
pulsation; Suitable for a variety of complex application combinations;
Full touch screen; A standard injector suitable for μ L specifications.
The vertical installation structure can be conveniently used in
conjunction with biological instruments such as micro manipulators and
stereotaxic brain locators, and is widely used in various fields of
biological experiments.
Functional Features
1. Operation mode: independent, reciprocating, synchronous, synchronous merging
2. Working mode: infusion, extraction, infusion/extraction, extraction/infusion, continuous, multiple infusion
3. Touch display: 5-inch high-definition touch screen, simple, clear, easy to understand, and controllable
4. Language types: Chinese and English, 2 languages to choose from
5. Online speed regulation: adjust the current flow rate without stopping the machine during operation
6.
Calibration function: Each channel injector can be independently
calibrated and calibration parameters can be stored independently
7.
Syringe selection: Built in from major domestic and foreign
manufacturers, with a choice of main models of syringes and support for
customization
8. Parameter settings: Each channel can independently
set liquid volume, flow rate, infusion time, extraction time, interval
time, and store them independently
9. Power on operation: Store the
current dynamic operating parameters when powered off, and resume
operation after powering on again
10. Reminder function: Real time
display of available Min/Max flow rates for each channel; Dynamically
display the operating parameters of each channel
11. Reminder
function: When each channel reaches its set target liquid volume, an
audible alarm will sound to remind each channel
12. Thrust setting: Each channel sets the appropriate thrust according to its own injector specifications
13. External control function: External level or pulse signal to achieve independent start stop control of each channel
14. Communication control: RS485/USB optional, touch settings for address and baud rate
15. Status output: Each channel independently outputs start stop and direction status
Technical Parameter
1. Number of channels: Four channels
2. Operation mode: independent, reciprocating, synchronous, synchronous merging
3. Working mode: infusion, extraction, infusion/extraction, extraction/infusion, continuous, multiple infusion
4. Maximum travel: 150mm
5. Accuracy:<1%
6. Repetitive accuracy: Error ≤ ± 0.05%
7. Sample injector specifications: 0.5 μ L to 60mL
8. Linear thrust: ≥ 160N
9. Linear speed range: 0.08 μ m/min to 160mm/min
10. Step resolution: 0.0625 μ m/μ step
11. Applicable power supply: AC 100V~240V 50/60Hz 30W
12. Working environment: Temperature: 0-40 ℃ Humidity: Humidity 20% -80% RH No condensation
13. Storage environment: Temperature: -10~60 ℃ Humidity: 20% -80% RH No condensation
14. Protection level: IP31
15. Product size (mm): 256 * 218 * 152 (mm)
Applicable Syringes
Optional syringe capacity | Inner diameter of syringe(mm) | Reference flow range(nL/min-mL/min) | Optional syringe capacity | Inner diameter of syringe(mm) | Reference flow range(nL/min-mL/min) |
0.5ul | 0.103 | 0.0007~0.0013 | 250ul | 2.304 | 0.3335~0.6671 |
1ul | 0.146 | 0.0013~0.0027 | 500ul | 3.256 | 0.6658~1.3322 |
2ul | 0.206 | 0.0027~0.0053 | 1ml |
4.73 |
1.4060~2.8110 |
5ul | 0.343 |
0.0074~0.0148 |
2ml |
8.92 |
4.9990~9.9980 |
10ul | 0.485 | 0.0148~0.0295 | 2.5ml |
8.92 |
4.9990~9.9980 |
25ul | 0.729 | 0.0334~0.0668 | 5ml |
12.4 |
9.6620~19.3220 |
50ul | 1.03 | 0.0667~0.1333 | 10ml |
14.87 |
13.8940~27.7860 |
100ul | 1.457 | 0.1334~0.2668 |
Ordering Information
Model | Name | Trip(mm) |
M5126 | Manual microinjection pump | 0-25 |
M5136 | Manual microinjection pump | 0-50 |
M5258 | High precision microinjection pump | 0-70 |
M5259 | Dual channel high-precision microinjection pump | 0-70 |
M5260 | Microinjection pump | 0-150 |
M5261 | Dual channel microinjection pump | 0-150 |
M5262 | Four channel microinjection pump | 0-150 |
M5263 | Ultra precision microinjection pump | 0-150 |
M5264 | Dual channel independent control microinjection pump | 0-150 |
Selection Guide
M5136 Manual microinjection pump |
M5258 High precision microinjection pump |
M5259 Dual channel high-precision microinjection pump |
M5260 Microinjection pump |
M5261 Dual channel microinjection pump |
M5262 Four channel microinjection pump |
M5263 Ultra precision microinjection pump |
M5264 Dual channel independent control microinjection pump |